ErgoX® Plus - ErgoTip Handle, Honey Brown | ErgoTip 1454-XP

ErgoDenta ApS · #1454-XP · Quelle

Herstellertext auf Englisch

The ErgoX® Plus ErgoTip handle in Honey Brown is a reusable double-ended handle for the ErgoTip system. Screw in your chosen tips, use them, then replace just the tip when it wears — so one handle works across many instruments. Lightweight, balanced and fully autoclavable.

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.

Step 1 - Pre-Cleaning

Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.

Step 2 - Automated Cleaning

Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.

Step 3 - Inspect & Maintain

Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.

Step 4 - Package

Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.

Step 5 - Sterilize

Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.

Step 6 - Store

Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.

Produktinformationen

HerstellerErgoDenta ApS
TypDouble-ended handle
FarbeHonigbraun
SystemErgoTip
MaterialNicht angegeben
AutoklavierbarJa
Sterilization Temperature134°C