IDactiv Instrument Disinfectant and Cleaner 1L Dosing Bottle

Topdental Products Ltd. · #BOS9842 · source

Bossklein IDactiv is a class IIb registered medical device, designed for the disinfection and cleaning of manual instruments and tools prior to terminal sterilisation using the method of immersion inside an instrument or ultrasonic bath. Alternatively, IDactiv can also be used to disinfect many types of rotating instruments such as burs, diamonds and polishers inside the bur bath.

Directions For Use

Add 20ml ± 1ml to 980ml of clean water for a 2% solution, 40ml ± 1ml to 960ml for a 4% solution. To achieve full efficacy use a 4% solution for a 60 minute soaking time. After disinfection, rinse instruments thoroughly with RO or distilled water prior to further processing and sterilisation. Follow the reprocessing instructions from the instrument manufacturer. Change the solution daily. For professional use only.

Testing & Efficacy

EN17111, EN14885, EN14561, EN14562, EN14563, EN14476, EN14348, EN13727, EN13624.

Effective Spectrum & Exposure Times

Micro-Organism: Bactericidal - Contact Time: 15 min - Concentration: 2% Pseudomonas aeruginosa - Contact Time: 15 min - Concentration: 2% Staphyloccocus aureus - Contact Time: 15 min - Concentration: 2% Enterococcus hirae - Contact Time: 15 min - Concentration: 2% E. coli - Contact Time: 15 min - Concentration: 2% Salmonella typhimurium - Contact Time: 15 min - Concentration: 2% MRSA - Contact Time: 15 min - Concentration: 2% Legionella pneumophila - Contact Time: 15 min - Concentration: 2% Yeasticidal: Candida albicans - Contact Time: 15 min - Concentration: 2% ENVELOPED VIRUSES: Vaccinia virus - Contact Time: 15 min - Concentration: 2% Hepatitis B Virus (HBV) - Contact Time: 15 min - Concentration: 2% Hepatitis C Virus (HCV) - Contact Time: 15 min - Concentration: 2% HIV - Contact Time: 15 min - Concentration: 2% Human T Cell Leukemia virus - Contact Time: 15 min - Concentration: 2% Measles virus - Contact Time: 15 min - Concentration: 2% Hepatitis Delta Virus (HDV) - Contact Time: 15 min - Concentration: 2% Influenza virus - Contact Time: 15 min - Concentration: 2% Coronavirus - Contact Time: 15 min - Concentration: 2% Herpesviridae - Contact Time: 15 min - Concentration: 2% Filoviridae (Ebola) - Contact Time: 15 min - Concentration: 2% Rubella virus - Contact Time: 15 min - Concentration: 2% Rabies virus - Contact Time: 15 min - Concentration: 2% Flavivirus - Contact Time: 15 min - Concentration: 2% Poxviridae - Contact Time: 15 min - Concentration: 2% NON-ENVELOPED VIRUSES: Adenovirus 5 - Contact Time: 30 min - Concentration: 4% Norovirus - Contact Time: 30 min - Concentration: 4% MYCOBACTERICIDAL: Mycobacterium avium - Contact Time: 60 min - Concentration: 4% TUBERCULOCIDAL: Mycobacterium terrae - Contact Time: 60 min - Concentration: 4%

Product Composition

Aqua, N-(3-Aminopropyl)-N-Dodecylpropane-1,3-Diamine (4.1% w/w 2372-82-9), DMPAP (3.5% w/w), Alcohols, C9-11, Ethoxylated, Tetrasodium N,N-Bis(Carboxylatomethyl)-L-Glutamate, Monoethanolamine, Corrosion Inhibitor, Perfume contains CINEOLE - Cineols - Eucolytus; HYDRATROPALDEHYDE; TRANS-MENTHONE; p-menth-1-en-4-ol., Dye.

Packing Details

6 x 1L Bottles per box

Physical Properties

Green liquid, Mild Fresh aroma, pH 9.00-10.00, Shelf life of 3 years

HTM 01-05 Compliance

3.3: Manual cleaning, governed by an appropriate protocol, is acceptable within the essential-quality-requirements framework. Within the best-practice framework, however, manual cleaning should be considered only where the manufacturer specifies that the device is not compatible with automated processes (including ultrasonic cleaning) or when the washer-disinfector is temporarily unavailable (for example for repair or validation). Exceptionally, where local experience indicates that pre-washing may be helpful (for example in the removal of tenacious dental materials), such action may be appropriate before automated cleaning.

3.4: New instruments should be cleaned and sterilized before using for the first time, unless supplied as sterile.

3.42: Effective cleaning of dental instruments before sterilization is of the utmost importance to reduce the risk of transmission of infectious agents.

3.5: Instruments cleaned as soon as possible after use may be more easily cleaned than those left for a number of hours before reprocessing.

3.6 When working with substances that can harden on instruments (for example cements), the instruments should be cleaned immediately. Instruments that cannot be cleaned should be discarded.

3.9: Manual cleaning should be considered where manufacturers' instructions specify that the device is not compatible with automated processes.

16.3: Always use detergents specifically formulated for manual cleaning of instruments.

Hazard CLP & Disposal

Danger. Causes skin irritation. Causes serious eye damage. Very toxic to aquatic life with long lasting effects. Wash hands thoroughly after handling. Wear protective gloves, eye protection, protective clothing. IF ON SKIN: Wash with plenty of water. IF IN EYES: Rinse cautiously with water for several minutes. Remove contact lenses, if present and easy to do. Continue rinsing. Immediately call a doctor. If skin irritation occurs: Get medical advice/attention. Collect spillage.

Storage. Store in a well-ventilated place. Keep container tightly closed. Keep cool. Contraindications. Heat Disposal. Do not discharge concentrate into drains or rivers.

UFI: 3X00-W0GP-900K-VWW7

UN: 1903, Packing group: III, Class: 8, Shipping name: DISINFECTANT, LIQUID, CORROSIVE, N.O.S. (N-(3-AMINOPROPYL)-N-DODECYLPROPANE-1,3-DIAMINE)

Product Information

ManufacturerTopdental Products Ltd.
pH9.00-10.00
AromaMild Fresh aroma
ClassClass IIb medical device
ColorGreen liquid
ShapeDosing Bottle
Efficacy2 in 1: cleans & disinfects
FeaturesIncludes corrosion inhibitor
Pack size1L
Packaging6 x 1L Bottles per box, 4 x 2.5L Bottles per box, 2 x 5L UN Approved Refill Bottles per box
En TestingEN17111, EN14885, EN14561, EN14562, EN14563, EN14476, EN14348, EN13727, EN13624
Shelf life3 years
Shelf Life3 years
ApplicationUse inside instrument bath or ultrasonic bath
Contact Time15 minutes
Product typeInstrument Disinfectant
Soaking Timefrom only 15 mins
Virus TestedEN14476
Safe For BursSafe to use on burs and rotating instruments
Dilution Ratio2% or 4%
Test StandardsEN 14885, EN 14561, EN 14562, EN 14563, EN 14476, EN 14348, EN 13727, EN 13624, ASTM E1052
TuberculocidalYes
Active Against TbYes
Concentration Options2% or 4% solution
Material CompatibilityWide material compatibility