Evostyle N.G.

New Life Radiology S.r.l. · source

Description

AC X-Ray Unit

9001M: Murale / Wall Mounted

9001S: Stativo/ Mobile

Why choose Evostyle N.G.?

  • Safe and easy to install. Simple control panel with LED buttonsand intuitive graphics. Two versions, the same utility: wall mounted (code 9001M) and mobile (code 9001S).
  • Exposure parameters adjustment. Manual setting available, to increase or decrease the exposure parameters value according to your needs. You can modify the preprogrammed exposure parameters to match the specifications of your image receptor (film, intraoral sensor or phosphor plate), the type of tooth (incisor, premolar, molar) or the patient morphology (adult or child).
  • Quality of details. Low dose setting for digital receptors. Automatic setting of exposure time. High heat capacity by combining an electric potential of 70 kVp and anodic current of 8 mA. Wide range of exposure times from 40ms to 2S.
  • Easy to install and to operate. The Evostyle N.G. is the simplest and easy to operate x-ray dental unit produced by New Life Radiology S.r.l. to deliver high image quality. Solid arm to reduce any risk of tube head vibration during the image acquisition. Three lengths arm, 40 cm, 80 cm, 110 cm. Hand switch with 3 m coiled cord.

Features

Classification

Class I Electro-medical Apparatus with type-B part appllied

Head

Self-rectifying single-phase enbloc assembly. Tube: KL2 -0.8-70G Toshiba DG-073B-AC

Type

70 kV

Rated electrical power

0,430 kW

Supply voltage

230/240 v (50/60 Hz) single - phase

Absorption

6 A

Focal spot

0,8 mm (IEC 336)

Anode current

8 mA

Total filtration

2 mm Al eq.

Fokus-skin distance

20 cm

Ø of long cone at the end of spacer

<6 cm

Dispersed radiation

≤ 0,25 mGy/h a 1meter from focus 1

Selectable anodic current

4-7 mA

Selectable voltage to x-ray tube

60-70 kVp

Intermittent operation

1 s exposure: 30 s reset max

Control logic

Deadman

Maximum apparent resistance of mains power supply

0,46 Ω

Fuse

Very Fast Acting F 6.3 A H

Alluminium filter to be interposed between x-ray window and collimator cone

AL, 1 mm (AIP99,9 UNI3567)

Deadman logic control

wireless

Informations produit

FabricantNew Life Radiology S.r.l.
FuseVery Fast Acting F 6.3 A H
HeadSelf-rectifying single-phase enbloc assembly. Tube: KL2 -0.8-70G Toshiba DG-073B-AC
TypeUnité radiologique AC
CategoryClass I Electro-medical Apparatus with type-B part appllied
Absorption6 A
Focal Spot0,8 mm (IEC 336)
Anode Current8 mA
Control LogicDeadman
Supply Voltage230/240 v (50/60 Hz) single - phase
Nominal Voltage70 kV
Total Filtration2 mm Al eq.
Alluminium FilterAL, 1 mm (AIP99,9 UNI3567)
Available VersionsWall mounted (9001M), Mobile (9001S)
Dispersed Radiation≤ 0,25 mGy/h a 1meter from focus 1
Exposure Time Range40ms to 2S
Fokus Skin Distance20 cm
Arm Lengths Available40 cm, 80 cm, 110 cm
Deadman Logic Controlwireless
Intermittent Operation1 s exposure: 30 s reset max
Rated Electrical Power0,430 kW
Hand Switch Cord Length3 m coiled cord
Selectable Anodic Current4-7 mA
Cone Diameter At End Of Spacer<6 cm
Selectable Voltage To X Ray Tube60-70 kVp
Maximum Apparent Resistance Of Mains Power Supply0,46 Ω