Howard 67 | ErgoDenta 7163 SP

ErgoDenta ApS · #7163 SP · source

Texte du fabricant en anglais

The Howard 67 harvests autogenous bone shavings for grafting.

  • Collects clean autogenous bone shavings
  • Ergonomic, controlled action
  • Reusable and fully autoclavable

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use. Step 1: Pre-Cleaning Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing. Step 2: Automated Cleaning Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry. Step 3: Inspect & Maintain Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant. Step 4: Package Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly. Step 5: Sterilize Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators. Step 6: Store Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.

Informations produit

FabricantErgoDenta ApS
TypeBone instrument
SterilisationEntièrement autoclavable