Wireless AC X-Ray Unit
8001M: Murale / Wall Mounted
8001S: Stativo / Mobile
Why choose Best -X-AC?
The radiographic unit BEST-X-AC is the fruit of the New Life Radiology long experience. It uses the most modern technique applied on dental radiology. It is light and compact, all functions integrated in the enbloc. Its lightness (Kg. 3,5) allows to operate the arm very easily. Through the exclusive timer placed in the enbloc, the exposition parameter set up with pre-set and adjustable software by the end user is made easier. The machine is remote controlled, with the operator's complete safety through the radio wave remote control. The exclusive thermoswitch system gives further safety by blocking Rx emissione in the event of a timer breakdown. BEST-X-AC is also avalable in Mobile Version on wheels.
Features
Classification: Class I Electro-medical Apparatus with type-B part appllied
Head: Self-rectifying single-phase enbloc assembly. Tube: KL2 -0.8-70G Toshiba DG-073B-AC
Type: 70 kV
Rated electrical power: 0,430 kW
Supply voltage: 230/240 v (50/60 Hz) single - phase
Absorption: 6 A
Focal spot: 0,8 mm (IEC 336)
Anode current: 8 mA
Total filtration: 2 mm Al eq.
Fokus-skin distance: 20 cm
Ø of long cone at the end of spacer: <6 cm
Dispersed radiation: ≤ 0,25 mGy/h a 1meter from focus 1
Selectable anodic current: 4-7 mA
Selectable voltage to x-ray tube: 60-70 kVp
Intermittent operation: 1 s exposure: 30 s reset max
Control logic: WIRELESS deadman
Maximum apparent resistance of mains power supply: 0,46 Ω
Fuse: Very Fast Acting F 6.3 A H
Alluminium filter to be interposed between x-ray window and collimator cone: AL, 1 mm (AIP99,9 UNI3567)
Deadman logic control: wireless
Product Information
| Manufacturer | New Life Radiology S.r.l. |
|---|---|
| Fuse | Very Fast Acting F 6.3 A H |
| Head | Self-rectifying single-phase enbloc assembly. Tube: KL2 -0.8-70G Toshiba DG-073B-AC |
| Type | 70 kV |
| Weight | 3,5 Kg |
| Category | Class I Electro-medical Apparatus with type-B part appllied |
| Absorption | 6 A |
| Focal Spot | 0,8 mm (IEC 336) |
| Anode Current | 8 mA |
| Control Logic | WIRELESS deadman |
| Supply Voltage | 230/240 v (50/60 Hz) single - phase |
| Total Filtration | 2 mm Al eq. |
| Alluminium Filter | AL, 1 mm (AIP99,9 UNI3567) |
| Available Versions | 8001M (Wall Mounted), 8001S (Mobile) |
| Dispersed Radiation | ≤ 0,25 mGy/h a 1meter from focus 1 |
| Fokus Skin Distance | 20 cm |
| Deadman Logic Control | wireless |
| Intermittent Operation | 1 s exposure: 30 s reset max |
| Rated Electrical Power | 0,430 kW |
| Selectable Anodic Current | 4-7 mA |
| Selectable Voltage To X Ray Tube | 60-70 kVp |
| Ø Of Long Cone At The End Of Spacer | <6 cm |
| Maximum Apparent Resistance Of Mains Power Supply | 0,46 Ω |